FDA recall record

Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit recall — FDA Z-0577-2015

It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold…

Class IIclassification
Z-0577-2015official record ID
Dec 17, 2014report date

Recall record at a glance

Official recordZ-0577-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmInvacare Corporation
Report dateDec 17, 2014
Recall initiation dateSep 23, 2014
Quantity5,963
LocationElyria, OH, United States

Official product description

Invacare Perfecto 2 V Oxygen Concentrator The finished device is packaged as a single unit. The service part contains a single pair of sieve beds; one pair of sieve beds is used in a single finished device. Intended to provide supplemental oxygen to patients with respiratory disorders, by separating nitrogen from room air, by way of a molecular sieve.

Why the product was recalled

It was determined that a defective sieve bed subassembly enabled zeolite sieve dust from the sieve beds to migrate into the valves of the manifold assembly and muffler causing a reduction in oxygen concentration.

Distribution information

Nationwide Distribution -- FL, NJ, GA, RI, PA, MI, CA, MA, IN, WA, CT, VA, NC, MO, AZ, CO, MS, MI, WI, TN, NH, IL, LA, OK, HI, KY, NE, SC, TX, IA, IN, and MS.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 1, 2022

Invacare TDX SP2, Model Nos recall — FDA Z-1130-2022

The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to…

Class IIFDA: OngoingInvacare Corporation