FDA recall record

Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use recall — FDA Z-0549-2019

Invacare has identified via customer complaints, the potential for the mounting bolt connecting the caster to the base frame of the patient lifts to…

Class IIclassification
Z-0549-2019official record ID
Dec 12, 2018report date

Recall record at a glance

Official recordZ-0549-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmInvacare Corporation
Report dateDec 12, 2018
Recall initiation dateNov 16, 2018
Quantity73,703 units
LocationElyria, OH, United States

Official product description

Invacare 9805 and 9805P Hydraulic Patient Lifts, Model Nos. 9805 and 9805P The intended use of a Non-AC-Powered patient lift is to lift and transfer a patient from one surface to another, as from a bed to a wheelchair. The device includes straps and a sling to support the patient.

Why the product was recalled

Invacare has identified via customer complaints, the potential for the mounting bolt connecting the caster to the base frame of the patient lifts to become loose. Loose hardware can cause wear to the hardware and housings. If left unresolved, the caster may separate from the base of the lift causing the Lift to tilt and a patient fall. Failure does not occur right away but happens over time if the lift is not properly maintained. Patient falls are inherently dangerous and may result in a range of injuries that are dependent on physical condition of the patient and the nature of the fall. When injuries occur, they are typically non-serious injuries such as bruising, lacerations, or abrasions. In rare cases, more serious injuries such as bone fracture, head trauma, or death can occur. For caregivers using the lift, attempts to manually stabilize a lift if it becomes unstable could result in sprains, serious back injuries, or bone fractures.

Distribution information

Worldwide Distribution: US (Nationwide) and countries of: Canada, Colombia, Germany, Guam, India, Northern Mariana Islands, Philippines and South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 1, 2022

Invacare TDX SP2, Model Nos recall — FDA Z-1130-2022

The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to…

Class IIFDA: OngoingInvacare Corporation