Recall record at a glance
| Official record | Z-1209-2015 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class I |
| FDA record status | Terminated |
| Recalling firm | Boston Scientific Corp |
| Report date | Mar 11, 2015 |
| Recall initiation date | Feb 3, 2015 |
| Quantity | 14 units |
| Location | San Jose, CA, United States |
Official product description
IntellaTip MiFi Open-Irrigated radiofrequency Ablation Catheters: Material number: M004EPM96200, Catalog Number: EPM9620: Material number: M004EPM9620K20, Catalog Number: EPM9620K2; Material number : M004EPM9620N40; Catalog number: EPM9620N4. The IntellaTip MiFi Open-Irrigated Ablation Catheter is a 7.5F (2.5 mm) quadrapolar open-irrigated (OI) ablation catheter designed to deliver radiofrequency (RF) energy to the 4.5 mm catheter tip electrode for cardiac ablation The product is not approved for use/sale in the US.
Why the product was recalled
Reports of formation of char adherent to the proximal part of the distal tip electrodes
Distribution information
Distributed in Germany, Great Britain, and the Netherlands.
How to read the classification
FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.