FDA recall record

Boston Scientific IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-10 recall — FDA Z-0465-2016

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate…

Class IIclassification
Z-0465-2016official record ID
Dec 30, 2015report date

Recall record at a glance

Official recordZ-0465-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBoston Scientific Corp
Report dateDec 30, 2015
Recall initiation dateNov 20, 2015
Quantity1556 total - all models
LocationSan Jose, CA, United States

Official product description

IntellaTip MiFi XP Asymmetric (N4) Curve 7/110/2.5/8-10 ; Material Number: M004PM4790N40; Catalog number: PM4790N4; Cardiac: The IntellaTip MiFi XP Catheter is indicated for interruption of accessory atrioventricular conduction pathways associated with tachycardia, for treatment of AV nodal reentrant tachycardia, for treatment of atrial flutter tachycardia, and for creation of complete AV block in patients with a rapid ventricular response to an atrial arrhythmia  typically chronic, drug refractory atrial fibrillation.

Why the product was recalled

During routine monitoring of complaints, BSC noted that the IntellaTip MiFi XP N4 complaint rate for distal tip bends/kinks was above the anticipated rate in BSC's product risk documentation.

Distribution information

Worldwide Distribution - US Nationwide in the state of: PA, FL, IN, MI, TN, AL, IL, KY, MI, TX, OH, MO, MA, MN, ID, AL, TN, VA, CA and internationally to Austria, Australia, Belgium, Switzerland, Germany, Great Britain, Greece, Hong Kong, Israel, Iran, Italy, Japan, Kuwait, Latvia, Malaysia, Netherlands, Saudi Arabia, Singapore, Thailand, and South Africa.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.