FDA recall record

Instrumentation Laboratory HemosIL Liquid Anti-Xa- Automated chromogenic assay for the recall — FDA Z-0149-2022

Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days

Class IIclassification
Z-0149-2022official record ID
Oct 27, 2021report date

Recall record at a glance

Official recordZ-0149-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmInstrumentation Laboratory
Report dateOct 27, 2021
Recall initiation dateAug 25, 2021
Quantity64,114
LocationBedford, MA, United States

Official product description

HemosIL Liquid Anti-Xa- Automated chromogenic assay for the quantitative determination of unfractionated heparin (UFH) and low molecular weight heparin (LMWH) activity in human citrated plasma on IL Coagulation Systems, (ACL TOP Family, ACL TOP Family 50 Seriest and ACL Elite/Elite Pro/8/9/10000) Part Number: 0020302600

Why the product was recalled

Labeled On-board instrument stability issue for current and future lots, reduced On-board Instrument Stability from 7 days to 5 days

Distribution information

US Nationwide Foreign: Canada Armenia Austria Australia Belgium Brunei Brazil Belarus Switzerland China Colombia Czech Republic Germany Denmark Spain Finland France United Kingdom Gibraltar Greece Hong Kong Hungary Indonesia Ireland Israel India Iran Italy¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿ Japan South Korea Kuwait Kazakhstan Lebanon Sri Lanka Lithuania Luxembourg Macau Mexico Malaysia Netherlands Poland Portugal Qatar Reunion Romania Serbia Russia Saudi Arabia Sweden Thailand Tunisia Turkey Taiwan Vietnam Zambia

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 5, 2026

GEM Premier 5000 recall — FDA Z-2825-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory
Medical DeviceAug 5, 2026

GEM Premier 5000 recall — FDA Z-2823-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory
Medical DeviceAug 5, 2026

GEM Premier 5000 recall — FDA Z-2824-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory
Medical DeviceAug 5, 2026

GEM Premier 7000 with iQM3 recall — FDA Z-2822-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory