FDA recall record

Instrumentation Laboratory GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 recall — FDA Z-2899-2026

The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing…

Class IIclassification
Z-2899-2026official record ID
Aug 12, 2026report date

Recall record at a glance

Official recordZ-2899-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmInstrumentation Laboratory
Report dateAug 12, 2026
Recall initiation dateMay 21, 2026
Quantity32 units
LocationBedford, MA, United States

Official product description

GEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010 00077430010 Product Description: The GEM Premier 7000 with iQM3 is a portable critical care system for use by health care professionals to rapidly analyze lithium heparinized whole blood samples at the point of health care delivery in a clinical setting and in a central laboratory.

Why the product was recalled

The active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunction could result in delayed recognition and management of Co-Oximetry and total bilirubin abnormalities and their sequelae.

Distribution information

Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MN, MO, MT, OH, PA, VA, WI, WV and the countries of China, Germany, Italy.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceAug 5, 2026

GEM Premier 5000 recall — FDA Z-2825-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory
Medical DeviceAug 5, 2026

GEM Premier 5000 recall — FDA Z-2823-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory
Medical DeviceAug 5, 2026

GEM Premier 5000 recall — FDA Z-2824-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory
Medical DeviceAug 5, 2026

GEM Premier 7000 with iQM3 recall — FDA Z-2822-2026

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Class IIFDA: OngoingInstrumentation Laboratory