FDA recall record

Stryker Neurovascular GDC-10 360 SOFT 10MM X 20CM SR Detachable Coil Sterile recall — FDA Z-1516-2016

The incorrect DFU was packaged with two lots of GDC 360 degree Detachable Coils.

Class IIclassification
Z-1516-2016official record ID
Apr 27, 2016report date

Recall record at a glance

Official recordZ-1516-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmStryker Neurovascular
Report dateApr 27, 2016
Recall initiation dateMar 21, 2016
Quantity316 total, both models
LocationFremont, CA, United States

Official product description

GDC-10 360 SOFT 10MM X 20CM SR Detachable Coil Sterile; Model number: M0033471020SR0; Neurology: GDC 360 Detachable Coils are intended for embolization of those intracranial aneurysms that because of their morphology, their location, or the patients general medical condition are considered by the treating neurosurgical team to be a) very high risk for management by traditional operative techniques, or b) inoperable.

Why the product was recalled

The incorrect DFU was packaged with two lots of GDC 360 degree Detachable Coils.

Distribution information

Worldwide Distribution: US (nationwide) including states of: OH, CO, WA, IL, MI, CA, TX and countries of: Hong Kong, Italy, Germany, France, Spain, Australia, Canada, Sweden, India, Turkey, Cyprus, Belgium, Switzerland, Czech Republic, Uruguay, Mexico, Saudi Arabia, Russia, Brazil, Netherlands, and Ukraine.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.