FDA recall record

Stryker Neurovascular The Synchro Neuro Guidewires are a steerable guidewire family with a recall — FDA Z-1939-2024

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of…

Class IIclassification
Z-1939-2024official record ID
Jun 12, 2024report date

Recall record at a glance

Official recordZ-1939-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmStryker Neurovascular
Report dateJun 12, 2024
Recall initiation dateApr 18, 2024
Quantity105,286
LocationFremont, CA, United States

Official product description

The Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of the Synchro 10 guidewire is 0.010in. The outside diameter of the Syncrho 14 guidewire is 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.

Why the product was recalled

Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

Distribution information

Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRAN, IRELAND, ISRAEL, ITALIA, JAPAN, JORDAN, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, NETHERLANDS, NORTHERN IRELAND, NORWAY, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UAE, UK.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.