FDA recall record

Abbott Medical Gallant HF Cardiac Resynchronization Therapy Defibrillator: HF CRT-D REF recall — FDA Z-2615-2023

Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D)…

Class IIclassification
Z-2615-2023official record ID
Sep 27, 2023report date

Recall record at a glance

Official recordZ-2615-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAbbott Medical
Report dateSep 27, 2023
Recall initiation dateAug 16, 2023
Quantity29,088 devices
LocationSylmar, CA, United States

Official product description

Gallant HF Cardiac Resynchronization Therapy Defibrillator: HF CRT-D REF CDHFA500Q Neutrino HF Cardiac Resynchronization Therapy Defibrillator: NxT HF CRT-D REF CDHFA600Q Entrant HF Cardiac Resynchronization Therapy Defibrillator : HF CRT-D REF CDHFA300Q The primary function of ICDs and CRT-Ds are to treat life-threatening ventricular tachyarrhythmias via anti-tachycardia pacing (ATP), or cardioversion/defibrillation. Implantable Cardioverter Defibrillators (ICDs) can either be a dual chamber (also known as DR if rate responsive) device that is connected to a right atrial (RA) pacing lead and right ventricular defibrillation lead or single chamber ICD (also known as VR if rate responsive) with only a defibrillation lead in the right ventricle (RV). The ICD, more generally known as an implantable pulse generator (IPG), is typically implanted in a subcutaneous device pocket located in the pectoral region and is attached to the pacing/defibrillator leads that are implanted through the transvenous technique with the distal tips positioned in the RA (dual chamber) and RV. Cardiac Resynchronization Therapy Defibrillator (CRT-D) systems consist of a right atrial (RA) pacing lead, a right ventricular (RV) defibrillation lead, as well as a third lead placed in the coronary venous system overlying the left ventricle (LV) to improve ventricular resynchronization through biventricular pacing for treatment of heart failure. CRT-D devices, more generally known as an implantable pulse generator (IPG), are typically implanted in a subcutaneous device pocket located in the pectoral region and are attached to the pacing/defibrillator leads that are implanted through transvenous technique.

Why the product was recalled

Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

Distribution information

U.S.: Nation-wide O.U.S.: Australia, Austria, Bahrain, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Nepal, Netherlands, New Zealand, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom. Latin America - Pending

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 19, 2025

Abbott Medical Eterna SCS IPG recall — FDA Z-0458-2026

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse…

Class IIFDA: OngoingAbbott Medical
Medical DeviceNov 19, 2025

Abbott Medical Liberta RC DBS IPG recall — FDA Z-0459-2026

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse…

Class IIFDA: OngoingAbbott Medical