FDA recall record

Abbott Medical The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model recall — FDA Z-0753-2025

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss…

Class IIclassification
Z-0753-2025official record ID
Jan 1, 2025report date

Recall record at a glance

Official recordZ-0753-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmAbbott Medical
Report dateJan 1, 2025
Recall initiation dateNov 21, 2024
Quantity7295 programmers
LocationSylmar, CA, United States

Official product description

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.

Why the product was recalled

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

Distribution information

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN TX, UT, VA, WA, WI, and WV. The countries of Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceNov 19, 2025

Abbott Medical Eterna SCS IPG recall — FDA Z-0458-2026

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse…

Class IIFDA: OngoingAbbott Medical
Medical DeviceNov 19, 2025

Abbott Medical Liberta RC DBS IPG recall — FDA Z-0459-2026

There is a potential issue affecting Eterna Spinal Cord Stimulation (SCS) Implantable Pulse Generators (IPGs), Model 32400 and Liberta RC Deep Brain Stimulation (DBS) Implantable Pulse…

Class IIFDA: OngoingAbbott Medical