FDA recall record

Tornier Fornier Phantom Fiber(TM) Sutures: Item Number Description SMB000425 4.5 MM PHANTOM recall — FDA Z-1374-2018

Data has shown that degradation of a raw material used in Phantom Fiber may occur more quickly than expected if improperly stored. This may impact the…

Class IIclassification
Z-1374-2018official record ID
Apr 25, 2018report date

Recall record at a glance

Official recordZ-1374-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmTornier, Inc.
Report dateApr 25, 2018
Recall initiation dateFeb 8, 2018
Quantity32253 units
LocationBloomington, MN, United States

Official product description

Fornier Phantom Fiber(TM) Sutures: Item Number Description SMB000425 4.5 MM PHANTOM FT BIOCOMPOSITE SUTURE ANCHOR WITH NEEDLES SMB000525 5.5 MM PHANTOM FT BIOCOMPOSITE SUTURE ANCHOR WITH NEEDLES SMP000425 4.5 MM PHANTOM FT PEEK SUTURE ANCHOR WITH NEEDLES SMP000525 5.5 MM PHANTOM FT PEEK SUTURE ANCHOR WITH NEEDLES SMSB0110 1PK PHANTOM FIBER SZ 2 WHITE SMSB0110N lPK BIOFIBER SUTURE SZ 2 WHT NEEDLED SMSB0112 10 PK PHANTOM FIBER SZ 2 WHITE SMSB0112N 10 PK PHANTOM FIBER SUTURE SZ 2WHT NEEDLED SMSB0120 1PK PHANTOM FIBER SZ 2 VIOLET SMSB0120N lPK BIOFIBER SUTURE SZ 2 VLT NEEDLED SMSB0122 10 PK PHANTOM FIBER SZ 2 VIOLET SMSB0122N 10 PK BIOFIBER SUTURE SZ 2 VLT NEEDLED SMSB0130 1PK PHANTOM FIBER SZ 2 WHITE/VIOLET SMSB0132 10 PK PHANTOM FIBER SZ 2 WHITE/VIOLET Indications for use for the Phantom Fiber are: Phantom Fiber suture is indicated for use in general soft tissue approximation and/or ligation, but not for use in cardiovascular or neurological tissues, microsurgery or ophthalmic surgery.

Why the product was recalled

Data has shown that degradation of a raw material used in Phantom Fiber may occur more quickly than expected if improperly stored. This may impact the shelf-life of the product which could cause the suture to lose strength more quickly than expected.

Distribution information

US Nationwide Distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceApr 15, 2026

Tornier HRS Max parts: 1 recall — FDA Z-1769-2026

Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.

Class IIFDA: OngoingTornier, Inc