FDA recall record

TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318 recall — FDA Z-1237-2026

A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during…

Class IIclassification
Z-1237-2026official record ID
Feb 11, 2026report date

Recall record at a glance

Official recordZ-1237-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmTornier, Inc.
Report dateFeb 11, 2026
Recall initiation dateDec 19, 2025
Quantity427
LocationBloomington, MN, United States

Official product description

TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318

Why the product was recalled

A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceApr 15, 2026

Tornier HRS Max parts: 1 recall — FDA Z-1769-2026

Affected products were incorrectly labeled as Tornier HRS Max products. These parts are compatible with the Tornier HRS system only, and are incompatible with the Tornier HRS Max system.

Class IIFDA: OngoingTornier, Inc