Getinge Usa Sales Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System recall — FDA Z-0422-2023

The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail…

Class IIclassification
Z-0422-2023official record ID
Dec 14, 2022report date

Recall record at a glance

Official recordZ-0422-2023
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmGetinge Usa Sales Inc.
Report dateDec 14, 2022
Recall initiation dateNov 7, 2022
Quantity65 US
LocationWayne, NJ, United States

Official product description

Flow-c Anesthesia System (Product Code 6887700) Flow-e Anesthesia System (Product Code 6887900)

Why the product was recalled

The power backup battery, under certain conditions, may exhibit reduced battery run-time while in use, and there is a potential that the battery may fail to provide the expected run time in the event of a mains power failure. This could require the clinician to use manual emergency ventilation of another ventilator is not available and the backup battery is exhausted.

Distribution information

US Nationwide distribution in the states of CO, FL, IN, MA, MO, and VA.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.