Recall record at a glance
| Official record | Z-0462-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Getinge Usa Sales Inc. |
| Report date | Nov 20, 2024 |
| Recall initiation date | Oct 3, 2024 |
| Quantity | 112 units |
| Location | Wayne, NJ, United States |
Official product description
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001B0
Why the product was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Distribution information
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.