Carestream Health DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243 recall — FDA Z-0800-2015

While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue…

Class IIclassification
Z-0800-2015official record ID
Dec 24, 2014report date

Recall record at a glance

Official recordZ-0800-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCarestream Health Inc.
Report dateDec 24, 2014
Recall initiation dateNov 25, 2014
Quantity629 units (Domestic: 358 units; Foreign: 271 units)
LocationRochester, NY, United States

Official product description

DRX-Revolution Mobile X-Ray System, MODEL DRXR-1, Service Code 1243, Catalog Number 1060177 (China) and 1019397 (all other countries) --- Carestream Health, Inc. The DRX-Revolution Mobile X-Ray System is a mobile imaging system that incorporates a self contained x-ray generator, image receptor, imaging display and software for acquiring medical diagnostic images outside of a standard x-ray room. The DRX-Revolution Mobile X-Ray System has a motor for driving the system so that the operator does not have to push the system, but can walk behind the system while it is moving from one location to another.

Why the product was recalled

While servicing a unit at a customer site, a field engineer discovered a broken bolt from the carriage assembly. Carestream has identified an issue related to the DRX-Revolution Mobile X-Ray System in which bolts that secure the x-ray tube assembly/boom to the device mounting block may break. It has been determined that this issue is a result of an improperly assembled part which holds the x-ray

Distribution information

Worldwide Distribution: US (nationwide) and countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, ITALY, JAPAN, KOREA, KUWAIT, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, and UNITED KINGDOM.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.