Recall record at a glance
| Official record | Z-1206-2023 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | Carestream Health, Inc. |
| Report date | Mar 8, 2023 |
| Recall initiation date | Jan 30, 2023 |
| Quantity | 72 units |
| Location | Rochester, NY, United States |
Official product description
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).
Why the product was recalled
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Distribution information
Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.