FDA recall record

Synthes DePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715 recall — FDA Z-0092-2015

Particle residue on the instrument from adhesive tape which was used to bind the device during transport. There may be a risk of tissue reaction if a…

Class IIclassification
Z-0092-2015official record ID
Oct 29, 2014report date

Recall record at a glance

Official recordZ-0092-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmSynthes, Inc.
Report dateOct 29, 2014
Recall initiation dateOct 2, 2014
Quantity11
LocationWest Chester, PA, United States

Official product description

DePuy Synthes Radiolucent Retractor, part number 387.580, lot number 3620715

Why the product was recalled

Particle residue on the instrument from adhesive tape which was used to bind the device during transport. There may be a risk of tissue reaction if a patient is exposed to latex particles from the adhesive tape, or a systemic reaction such as anaphylaxis in patients with latex allergy.

Distribution information

US Distribution including the states of PA, NC, TN, NY, WI, WA and MI.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.