Recall record at a glance
| Official record | Z-2698-2020 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Synthes, Inc. |
| Report date | Aug 5, 2020 |
| Recall initiation date | Jun 18, 2020 |
| Quantity | 104 units |
| Location | West Chester, PA, United States |
Official product description
MatrixNEURO Screws - Product Usage: The intended use for this device is in fixation of cranial bones and facial skeleton in procedures such as craniotomies, reconstructive procedures, fracture repair, osteotomies, and selective orthognathic surgery of the maxilla and chin.
Why the product was recalled
One lot of MatrixNEURO 5mm screws was packaged in a body clip etched "4", which indicated a 4mm screw. The outer package labeling correctly identifies a 5mm screw, but the body clop incorrectly identifies a shorter length for the screw.
Distribution information
International distribution including in the countries of Australia, Belgium, Chile, France, Slovakia, South Korea.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.