FDA recall record

Intuitive Surgical daVinci Harmonic ACE Curved Shears recall — FDA Z-0841-2019

During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The…

Class IIclassification
Z-0841-2019official record ID
Feb 27, 2019report date

Recall record at a glance

Official recordZ-0841-2019
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmIntuitive Surgical, Inc.
Report dateFeb 27, 2019
Recall initiation dateJan 16, 2019
Quantity43755 inserts
LocationSunnyvale, CA, United States

Official product description

daVinci Harmonic ACE Curved Shears

Why the product was recalled

During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less than a few millimeters) on the pouch leading to a potential breach in sterility.

Distribution information

US and Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kuwait, Lebanon, Mexico, Monaco, Netherlands, Pakistan, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Korea, Spain, Taiwan, Turkey, United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Da Vinci 5 ASSY,DV5 TOWER,IS5000 recall — FDA Z-2310-2026

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the…

Class IIFDA: OngoingIntuitive Surgical, Inc.