FDA recall record

Intuitive Surgical da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi recall — FDA Z-0999-2022

As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally…

Class IIclassification
Z-0999-2022official record ID
May 4, 2022report date

Recall record at a glance

Official recordZ-0999-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmIntuitive Surgical, Inc.
Report dateMay 4, 2022
Recall initiation dateMar 17, 2022
Quantity6,8881 boxes (or 413,286 individual units/6 instruments/box)
LocationSunnyvale, CA, United States

Official product description

da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45 Curved-Tip REF 480545 da Vinici X da Vinci Xi Sureform 60 REF 480460

Why the product was recalled

As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected. Overall rate of harm associated with bleeding of critical severity exceeds the predicted likelihood of harm as documented within Intuitive s existing risk documentation.

Distribution information

Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. The countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Monaco, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Da Vinci 5 ASSY,DV5 TOWER,IS5000 recall — FDA Z-2310-2026

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the…

Class IIFDA: OngoingIntuitive Surgical, Inc.