FDA recall record

Intuitive Surgical da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11 recall — FDA Z-0702-2022

Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip.…

Class IIclassification
Z-0702-2022official record ID
Mar 9, 2022report date

Recall record at a glance

Official recordZ-0702-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmIntuitive Surgical, Inc.
Report dateMar 9, 2022
Recall initiation dateDec 23, 2021
Quantity149,646 reducers
LocationSunnyvale, CA, United States

Official product description

da Vinci Xi and X 12-8 mm Conductive Cannula Reducer, REF: 470381-11

Why the product was recalled

Conductive Cannula Reducer metal tip may get dislodged from the plastic shaft. If unnoticed, additional surgery may be required to remove the metal tip. If the procedure proceeds without the metal tip, there is potential for energy leakage, which could cause tissue damage. If the metal tip remains in the patient, adhesions may form, which may lead to bowel obstruction and chronic abdominal pain.

Distribution information

Worldwide distribution - US Nationwide distribution in the states of California, Wisconsin, Virginia, Ohio, Texas, New York, Nevada, Connecticut, New Jersey, Massachusetts, Georgia, Florida, Indiana, North Carolina, Missouri, Pennsylvania, Michigan, Oregon, Kentucky, Hawaii, Tennessee, Alabama, Nebraska, Maryland, Oklahoma, Colorado, Arkansas, Arizona, Kansas, Illinois, South Carolina, Minnesota, North Dakota, Iowa, Louisiana, Washington, Delaware, Mississippi, New Mexico, District of Columbia, Idaho, New Hampshire, Rhode Island, West Virginia, South Dakota, Alaska, Maine, Utah, Montana, Vermont, Wyoming, Puerto Rico and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Luxembourg, Martinique, Monaco, Netherlands, Norway, Poland, Qatar, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Da Vinci 5 ASSY,DV5 TOWER,IS5000 recall — FDA Z-2310-2026

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the…

Class IIFDA: OngoingIntuitive Surgical, Inc.