da Vinci X, Xi Surgical System, & da Vinci 5 Surgical System recall — FDA Z-1232-2026

Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was…

Class IIclassification
Z-1232-2026official record ID
Feb 11, 2026report date

Recall record at a glance

Official recordZ-1232-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmIntuitive Surgical, Inc.
Report dateFeb 11, 2026
Recall initiation dateNov 7, 2025
Quantity219
LocationSunnyvale, CA, United States

Official product description

Brand Name: Da Vinci Product Name: da Vinci X, Xi Surgical System, & da Vinci 5 Surgical System Model/Catalog Number: IS4000; IS4200; IS5000 Software Version: N/A Product Description: The da Vinci Xi, X, and da Vinci 5 surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument

Why the product was recalled

Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction

Distribution information

U.S.: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. O.U.S.: Austria, Brazil, China, France, Germany, Greece, India, Italy, Japan, Lebanon, Netherlands, South Korea, Spain, Switzerland, Taiwan and United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceJun 10, 2026

Da Vinci 5 ASSY,DV5 TOWER,IS5000 recall — FDA Z-2310-2026

Due to unintentionally activated near infrared handheld camera when attempting to turn off illumination during normal clinical workflow transitions, particularly when switching from the…

Class IIFDA: OngoingIntuitive Surgical, Inc.