FDA recall record

Accuray CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be recall — FDA Z-1023-2025

Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right…

Class IIclassification
Z-1023-2025official record ID
Feb 5, 2025report date

Recall record at a glance

Official recordZ-1023-2025
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusCompleted
Recalling firmAccuray Inc.
Report dateFeb 5, 2025
Recall initiation dateDec 23, 2024
Quantity212
LocationMadison, WI, United States

Official product description

CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery System is indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated.

Why the product was recalled

Snap rings may become partially or completed detached from the shafts within the internal mechanism. Uncontrolled rotation may occur on the left and right roll axes when the rings are removed or damaged.

Distribution information

Worldwide - US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records