FDA recall record

Accuray CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided recall — FDA Z-1272-2021

The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.

Class IIclassification
Z-1272-2021official record ID
Mar 24, 2021report date

Recall record at a glance

Official recordZ-1272-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmAccuray Inc.
Report dateMar 24, 2021
Recall initiation dateFeb 26, 2021
Quantity243 units
LocationMadison, WI, United States

Official product description

CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Why the product was recalled

The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.

Distribution information

Worldwide distribution - US Nationwide distribution including in the states of AK, AZ. CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV, PR and the countries of Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Republic of Korea, Latvia, Luxembourg, Malaysia, Mexico, Oman, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.