FDA recall record

ARJOHUNTLEIGH POLSKA Sp. z.o.o Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 recall — FDA Z-1452-2020

The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain…

Class IIclassification
Z-1452-2020official record ID
Mar 18, 2020report date

Recall record at a glance

Official recordZ-1452-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmARJOHUNTLEIGH POLSKA Sp. z.o.o.
Report dateMar 18, 2020
Recall initiation dateAug 22, 2019
Quantity669 trollys
LocationKomorniki, Poland

Official product description

Concerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB5000-01 [UDI: (01)05055982788144]. and Concerto & Basic Shower Trolley Spare parts (safety catch) part no. 8451622, 8546608, S8533571-031, S8533572-031, S8533570-031 S8542201-014, S8523559-031,

Why the product was recalled

The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall. The patient falling could lead to various types of injuries.

Distribution information

US: CA, CO, CT, GA, MA, MD, MI, NC, NJ, NY, OH, OK, TX, VA, WI OUS: Australia Austria, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iceland, India, Ireland, Italy, Japan, Korea, Luxembourg, Netherlands, Netherlands/CW, New Zealand, Norway, Philippines, Poland, Romaina, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province of China, United Arab Emirates, United Kingdom

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.