FDA recall record

ARJOHUNTLEIGH POLSKA Sp. z.o.o Arjo Citadel beds originally assembled with the IndiGo recall — FDA Z-2588-2021

The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission…

Class IIclassification
Z-2588-2021official record ID
Oct 6, 2021report date

Recall record at a glance

Official recordZ-2588-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmARJOHUNTLEIGH POLSKA Sp. z.o.o.
Report dateOct 6, 2021
Recall initiation dateJun 10, 2021
Quantity367 units
LocationKomorniki, N/A, Poland

Official product description

Arjo Citadel beds originally assembled with the IndiGo modules during manufcturing, model number CX812A3J3AMAD0; and IndiGo modules (retrofit kits) that can be installed on Citadel (model no. INDIDAA) or Enterprise (model no. INDIJAA) beds manufactured without IndiGo modules assembled at the time of manufacturing and subsequently retrofitted on the beds.

Why the product was recalled

The IndiGo power cord might wear during use which could lead to its damage, resulting in the cord being cut, burnt or short circuit with limited emission of sparks.

Distribution information

US Nationwide distribution in the states of CA, FL, IL, KY, MD, NV, NY, PA, and SD.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records