FDA recall record

CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed recall — FDA Z-2449-2015

Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching…

Class IIclassification
Z-2449-2015official record ID
Sep 2, 2015report date

Recall record at a glance

Official recordZ-2449-2015
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCarestream Health Inc.
Report dateSep 2, 2015
Recall initiation dateAug 7, 2015
QuantityDomestic: 76 units, Foreign: 160 units
LocationRochester, NY, United States

Official product description

CARESTREAM DRX-EVOLUTION X-Ray System Product Usage: The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.

Why the product was recalled

Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.

Distribution information

Worldwide Distribution - US Nationwide and in the countries: Australia, Canada, Chile, China, India, Korea, Malaysia, New Zealand, Singapore, Switzerland, Taiwan and Thailand.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records