BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system recall — FDA Z-0997-2026

Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

Class IIclassification
Z-0997-2026official record ID
Dec 31, 2025report date

Recall record at a glance

Official recordZ-0997-2026
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmBeckman Coulter Ireland, Inc.
Report dateDec 31, 2025
Recall initiation dateOct 14, 2025
Quantity36,981 reagents (U.S.-35,320, O.U.S.-1,661)
LocationO'Callaghan'S Mills, N/A, Ireland

Official product description

BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers.

Why the product was recalled

Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

Distribution information

Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guan, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Brazil, Canada, Ecuador, Ireland, Mexico, Netherlands, Panama, Taiwan, and Province of China.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.