Recall record at a glance
| Official record | Z-1815-2025 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class III |
| FDA record status | Ongoing |
| Recalling firm | Beckman Coulter Ireland, Inc. |
| Report date | Jun 4, 2025 |
| Recall initiation date | Apr 23, 2025 |
| Quantity | 2747 units |
| Location | O'Callaghan'S Mills, N/A, Ireland |
Official product description
Apo B Reagent, REF: OSR6143
Why the product was recalled
The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.
Distribution information
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.
How to read the classification
FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.