FDA recall record

Cytocell Aquarius LPH 511-A CRLF2 Distal Probe Green, model no recall — FDA Z-1099-2021

Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a…

Class IIclassification
Z-1099-2021official record ID
Feb 17, 2021report date

Recall record at a glance

Official recordZ-1099-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCytocell Ltd.
Report dateFeb 17, 2021
Recall initiation dateDec 22, 2020
Quantity72
LocationCambridge, N/A, United Kingdom

Official product description

Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene

Why the product was recalled

Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.

Distribution information

Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.