FDA recall record

Cytocell Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in recall — FDA Z-0741-2022

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The…

Class IIclassification
Z-0741-2022official record ID
Mar 16, 2022report date

Recall record at a glance

Official recordZ-0741-2022
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCytocell Ltd.
Report dateMar 16, 2022
Recall initiation dateJan 18, 2022
Quantity19 kits
LocationCambridge, N/A, United Kingdom

Official product description

Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024

Why the product was recalled

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

Distribution information

US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.