FDA recall record

3M Company - Health Care Business 3M Bair Hugger(TM) Normothermia System, Temperature recall — FDA Z-0114-2018

During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a…

Class IIclassification
Z-0114-2018official record ID
Nov 29, 2017report date

Recall record at a glance

Official recordZ-0114-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firm3M Company - Health Care Business
Report dateNov 29, 2017
Recall initiation dateOct 17, 2017
Quantity153 cases (3825 sensors)
LocationSaint Paul, MN, United States

Official product description

3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)

Why the product was recalled

During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.

Distribution information

CA, MI, IL, MO, MN, GA, and Canada, Switzerland

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.