Recall record at a glance
| Official record | Z-1475-2024 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Ongoing |
| Recalling firm | 3M Company - Health Care Business |
| Report date | Apr 10, 2024 |
| Recall initiation date | Feb 20, 2024 |
| Quantity | 265,435 eaches |
| Location | Saint Paul, MN, United States |
Official product description
3M Tube Securement Device, Small, 1.7 IN/PO x 2.4 IN/PO, REF 1500U, Sterile; and 3M Tube Securement Device, Large, 2.2 IN/PO x 3.5 IN/PO, REF 1501U, Sterile. Product is packaged 1 device per pouch, 25 pouches per primary carton, and 4 primary cartons per shipper box.
Why the product was recalled
Manufacturing nonconformities are in some lots of the Tube Securement Device, such as the top film layer may be partially or completely missing and some devices may exhibit the appearance of an incorrect shape due to excess material not being removed from the product liner.
Distribution information
Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MN, NC, NH, NY, OH, PA, TX, VA, and WA, including PR. The countries of Australia, Austria, Canada, China, Finland, Germany, Ireland, New Zealand, Norway, Poland, Sweden, and the United Kingdom.
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.