FDA recall record

Mizuho OSI 216 Wilson Frame Pads, Model # 5319-37 recall — FDA Z-1867-2017

Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient…

Class IIclassification
Z-1867-2017official record ID
Apr 26, 2017report date

Recall record at a glance

Official recordZ-1867-2017
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmMizuho OSI
Report dateApr 26, 2017
Recall initiation dateMar 28, 2017
Quantity216 sets
LocationUnion City, CA, United States

Official product description

216 Wilson Frame Pads, Model # 5319-37; Used with Radiolucent Wilson Frames, model numbers 5319G and 5321G. General and Plastic Surgery: These pads are used as a component of the Wilson Frames, which support the torso of a prone patient during spine surgery. The pads are supplied with the frames and as replacements.

Why the product was recalled

Mizuho has discovered that the 5319-37 Wilson Frame Pads have a design deficiency in the construction of their mounting sleeves that may allow the patient to contact the underlying carbon fiber structure of the Wilson frame and could lead to a pressure injury.

Distribution information

United States, Canada, Japan, Australia, New Zealand, Great Britain, Sweden, Spain, Germany, Switzerland, Kenya, Saudi Arabia, Hong Kong, South Korea.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceFeb 17, 2016

Mizuho OSI Trios, Table Base recall — FDA Z-0792-2016

Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or…

Class IIFDA: TerminatedMizuho OSI