FDA recall record

Mizuho CE IPX4, mode MOT-5602BW, 100-240, 50-60Hz, 400VA, REF 18-140-20 containing hand recall — FDA Z-1293-2021

There is a potential that hand control units may cause the tabletop to slide in the opposite direction from that selected by the user. This could result…

Class IIclassification
Z-1293-2021official record ID
Mar 31, 2021report date

Recall record at a glance

Official recordZ-1293-2021
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmMizuho OSI
Report dateMar 31, 2021
Recall initiation dateFeb 8, 2021
Quantity26 units
LocationUnion City, CA, United States

Official product description

Mizuho CE IPX4, mode MOT-5602BW, 100-240, 50-60Hz, 400VA, REF 18-140-20 containing hand control Unit - Product Usage: intended to support a patient during surgical operations.

Why the product was recalled

There is a potential that hand control units may cause the tabletop to slide in the opposite direction from that selected by the user. This could result in a delay in procedure.

Distribution information

Worldwide distribution - US Nationwide distribution in the state of CA, and the countries of Costa Rica, Chile, Bolivia, Peru.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.