Cepheid Xpert SA Nasal Complete recall — FDA Z-0414-2026
As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type.
Monitor U.S. medical device recall enforcement reports from the FDA. Review product descriptions, recall reasons, classifications, dates, recalling firms and distribution information where the FDA provides them.
21,444 detailed records are currently available in this category.
As a result of Post-Market half-time stability testing, false negative test result was observed for a positive sample type.
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary…
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual…
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Due to incomplete seals in the pouch which provide the sterile barrier.
The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots…
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient…
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage…
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Potential for contamination with cotton fibers.
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
It was determined there is the potential of silicone fragments detaching from the yellow sealing component.
It was determined there is the potential of silicone fragments detaching from the yellow sealing component.
Due to incomplete seals in the pouch which provide the sterile barrier.
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
Medline has identified a limited quantity of saline IV flush syringes that may not meet the required concentration of sodium chloride.
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly…
Medical device recall records can change as the FDA receives additional information. Check the official record identifier and current FDA information before making health or product-use decisions.
Recall Lookup USA keeps the last complete source record available when an official data service is temporarily unavailable. When a source provides only a few fields, we show only those known facts rather than padding the page with generic text.
This database is designed for discovery and reference. It can help you locate a relevant recall record, identify the recalling firm or NHTSA campaign, compare dates and classifications, and follow the official source. It is not a substitute for a VIN-specific NHTSA check, a manufacturer notice, medical advice or instructions issued by the recalling company.