Arthrex BioSync Anatomic Cotton Wedge, 16 x 6.5 mm recall — FDA Z-1960-2016
Potential for polystyrene particulate presence.
Monitor U.S. medical device recall enforcement reports from the FDA. Review product descriptions, recall reasons, classifications, dates, recalling firms and distribution information where the FDA provides them.
28,992 detailed records are currently available in this category.
Potential for polystyrene particulate presence.
Potential for polystyrene particulate presence.
The firm became aware that when using EtC02 in the kPa or % setting and in a situation where the reading is above 9.9 kPa, the display of the LIFEPAK 15 respiratory rate may partially…
The firm is adding the warning statement to the Instruction for Use for the neoBLUE LED Phototherapy System and provide a laminated Quick Guide to affix to each device.
Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the…
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may…
Complaints of smoke coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. In…
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the…
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the…
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may…
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
Zimmer Biomet is initiating a recall of specific lots of foil pouched, sterile Trabecular Metal implants that were manufactured between April 2011 and October 2012. The foil pouches are…
Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the…
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
The device is reported to enter safe mode and is no longer able to charge batteries.
Potential for polystyrene particulate presence.
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the…
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may…
There is a potential for elevated hemoglobin (Hgb) and impedance platelet count (PLT-I) values.
During an FDA inspection it was found that the products are marketed without a cleared 510k, nor is approved for multiple use.
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the…
Zimmer Biomet is initiating a recall of specific lots of foil pouched, sterile Trabecular Metal implants that were manufactured between April 2011 and October 2012. The foil pouches are…
PICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the…
Medical device recall records can change as the FDA receives additional information. Check the official record identifier and current FDA information before making health or product-use decisions.
Recall Lookup USA keeps the last complete source record available when an official data service is temporarily unavailable. When a source provides only a few fields, we show only those known facts rather than padding the page with generic text.
This database is designed for discovery and reference. It can help you locate a relevant recall record, identify the recalling firm or NHTSA campaign, compare dates and classifications, and follow the official source. It is not a substitute for a VIN-specific NHTSA check, a manufacturer notice, medical advice or instructions issued by the recalling company.