FDA recall record

Stryker PowerPRO (XT, IT, or TL) Cots Stretcher, Wheeled The Stryker Power-PRO" are powered recall — FDA Z-1990-2016

Complaints of smoke coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after…

Class IIclassification
Z-1990-2016official record ID
Jun 22, 2016report date

Recall record at a glance

Official recordZ-1990-2016
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmStryker Medical Division of Stryker Corporation
Report dateJun 22, 2016
Recall initiation dateMay 5, 2016
Quantity770
LocationPortage, MI, United States

Official product description

Stryker PowerPRO (XT, IT, or TL) Cots Stretcher, Wheeled The Stryker Power-PRO" are powered ambulance cots are designed to support and transport a maximum weight of 700 lb. (318 kg) in pre-hospital and hospital environments.

Why the product was recalled

Complaints of smoke coming from the foot end of the cot. This reportedly occurred when either the battery was placed into the cot or shortly after placing the battery into the cot. In addition to the alleged smoking, one of the reports had also noted that when the user attempted to lower the cot (prior to the alleged smoking) the device went into high speed retract and then a few moments later i

Distribution information

Distributed US (nationwide) and the countries of Australia, Canada, Hong Kong, Spain, Switzerland, and United Kingdom.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records