Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX…
Update to the carryover avoidance technical bulletin to detail that IgA and IgM assays should not be run immediately after Fructosamine Catalogue Number FR3133 or FR4030 on the RX…
***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for…
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration…
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration…
Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than…
Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient…
When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range
Standard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples
Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been…
Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine
Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional…
An update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Lactate Dehydrogenase assay on RX Instruments. Additional…
Amylase AY3805 batch 480449 and Pancreatic Amylase AY3855 batch 480483 have failed to meet the quoted performance claims prior to expiry on 28th Sept 2020.
Amylase AY3805 batch 480449 and Pancreatic Amylase AY3855 batch 480483 have failed to meet the quoted performance claims prior to expiry on 28th Sept 2020.
The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly.
Class IIFDA: OngoingRandox Laboratories Ltd.
Understanding this recall history
This page summarizes recall records in which the official source identifies Randox Laboratories Ltd. as the recalling firm or manufacturer. The number of records should not be treated as a safety rating or a measure of company size, because recall counts depend on product mix, reporting periods and how government records are organized.