FDA recall record

Randox Laboratories CK MB Calibrator, Catalogue Number CK2393 recall — FDA Z-1097-2022

Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with…

Class IIIclassification
Z-1097-2022official record ID
May 25, 2022report date

Recall record at a glance

Official recordZ-1097-2022
CategoryMedical Device Recalls
ClassificationClass III
FDA record statusTerminated
Recalling firmRandox Laboratories Ltd.
Report dateMay 25, 2022
Recall initiation dateMar 25, 2022
Quantity1,392
LocationCrumlin (North), N/A, Ireland

Official product description

CK MB Calibrator, Catalogue Number CK2393

Why the product was recalled

Randox have realigned CK-MB Calibrator, CK2393, lot number 4410CK. Calibrating with the new value will cause a positive shift of approximately 6% with both Quality Controls and Patient Samples.

Distribution information

Domestic distribution to WV and PR. International distribution to Bangladesh, Bosnia & Herzegovina, Botswana, China, Colombia, Costa Rica, Croatia, Egypt, Fiji, France, Georgia, Germany, Ghana, Greece, Grenada, Hong Kong, India, Iraq, Italy, Lithuania, Malaysia, Maldives, Malta, Mexico, Morocco, Mozambique, Nepal, Nigeria, Pakistan, Philippines, Poland, Puerto Rico, Romania, Russia, Saudi Arabia, South Korea, Switzerland, Thailand, Turkey, UK, Uruguay, Uzbekistan and Vietnam.

How to read the classification

FDA Class III generally covers situations in which use of or exposure to the product is not likely to cause adverse health consequences. The classification displayed here is taken from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.