Coltene Whaledent Strauss 25 Carbide Dental Burs (100057XXV) recall — FDA Z-0468-2021
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.
Complete FDA and NHTSA recall records dated in November 2020.
These records include useful source details for the selected month.
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.
Tightening the LIGAPASS Band with this instrument, the tensioner's pliers can crush the band causing the band to break during surgery and lead to a risk of infection and re-intervention ,
Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than…
Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures
Some electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not…
Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the…
Potential Incorrect Validation of Results Due to an Erroneous QC Status When Using Status Expiration Control Rule
The firm is replacing all tubes that were manufactured and distributed prior to Enforcement Discretion Approval.
The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in…
The Instructions for Use (IFU) updated 03/02/2020 does not match data listed in the post-approval study.
Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROS…
The feeding spike sets may leak at the interface of the tube and spike connector.
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.
The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in…
The flocked tip of the 100mm swab may break off in the nasophyaryngeal cavity during sample collection. This could result in soft tissue injury of the nasophyaryngeal cavity or a delay in…
Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROS…
The feeding spike sets may leak at the interface of the tube and spike connector.
It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm…
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.
The device has the potential to release nickel above the acceptable margin of safety in pediatric patients who weigh less than 20 kg (44.09 lbs) when 2 or more implants are used.
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.
Due to an extra digit being inadvertently added to the serial number of a subset of implantable pulse generators, they become not compatible with therapy programmers with software version…
FDA has informed Micro-Tech that the Elastic Traction Device is not appropriately classified as a Class I device and requires 510(k) clearance to be marketed in the US.