This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of…
There is a potential for some systems built between 2014 and Jan 2018 to have a light output falling below the recommended minimum output. This problem has been associated with exposure of…
Inability to dose the test strips, may result in a temporary inability to get a result that could lead to a delay in treatment and/or therapy decisions
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C…
Potential for thermal damage near the power cord connector and Power Entry Module (PEM) due to improper connection of the power cord connector and PEM pins on the Cocoon Convective Warming…
Potential for thermal damage near the power cord connector and Power Entry Module (PEM) due to improper connection of the power cord connector and PEM pins on the Cocoon Convective Warming…
Fetal transducers distributed in advance of receiving 510(k) and international clearances. Possible adverse clinical impact: inability to accurately detect and measure fetal heart rate,…
Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were…
The gowns contain an unknown level of bioburden, unknown microbiological flora, and unknown particulate contamination.