Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol…
Baxter r recalled this device due to the use of an incorrect plastic formulation in the manufacture of the blue luer component of the FLOSEAL malleable tips.
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken…
Class IIFDA: TerminatedMedical Components, Inc dba MedComp
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken…
Class IIFDA: TerminatedMedical Components, Inc dba MedComp
Sample material may potentially come in contact with the pipetting nozzle during operation of cobas p 612 pre-analytical system (63x) (LCP1) with non-filter tips. The possible presence of…
Class IIFDA: TerminatedRoche Diagnostics Corporation
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken…
Class IIFDA: TerminatedMedical Components, Inc dba MedComp
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken…
Class IIFDA: TerminatedMedical Components, Inc dba MedComp
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken…
Class IIFDA: TerminatedMedical Components, Inc dba MedComp
OrthoPAT reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means…
Elevated false positive results Yersinia enterocolitica (Yersinia) from customers using the EP Nucleic Acid Test Cartridges component (20-006-023) from the Verigene EP Nucleic Acid Test Kit…
Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems…
Class IIFDA: TerminatedSiemens Healthcare Diagnostics, Inc.
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken…
Class IIFDA: TerminatedMedical Components, Inc dba MedComp