FDA recall record

Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps recall — FDA Z-2263-2024

Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the…

Class IIclassification
Z-2263-2024official record ID
Jul 10, 2024report date

Recall record at a glance

Official recordZ-2263-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmZyno Medical LLC
Report dateJul 10, 2024
Recall initiation dateJun 4, 2024
Quantity88 units
LocationNatick, MA, United States

Official product description

Zyno Medical Z800, Z800F, Z800W, and Z800WF Infusion Pumps

Why the product was recalled

Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.

Distribution information

US Nationwide distribution.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.