FDA recall record

ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator recall — FDA Z-2324-2024

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during…

Class IIclassification
Z-2324-2024official record ID
Jul 17, 2024report date

Recall record at a glance

Official recordZ-2324-2024
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusOngoing
Recalling firmZoll Medical Corporation
Report dateJul 17, 2024
Recall initiation dateMay 31, 2024
Quantity14885
LocationDeerfield, WI, United States

Official product description

ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C

Why the product was recalled

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest

Distribution information

Nationwide including PR. Foreign: AU BM CA CH DE DK ES FR GB HK HU ID IE IL IT KR MA MT NL NO PR SA SE SI SN TR

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.

Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1321-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1319-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1320-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1322-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1316-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation
Medical DeviceMar 19, 2025

ZOLL Powerheart G5 AED recall — FDA Z-1317-2025

The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).

Class IIFDA: OngoingZOLL Medical Corporation