FDA recall record

Biomet Zimmer Biomet Comprehensive Segmental Revision System (SRS) recall — FDA Z-1173-2018

During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A…

Class IIclassification
Z-1173-2018official record ID
Apr 4, 2018report date

Recall record at a glance

Official recordZ-1173-2018
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmBiomet, Inc.
Report dateApr 4, 2018
Recall initiation dateFeb 28, 2018
Quantity634
LocationWarsaw, IN, United States

Official product description

Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral Reconstruction, and Total Humeral Reconstruction

Why the product was recalled

During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.

Distribution information

Distribution US Nationwide

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.