Cook Ireland Zilver 518 Vascular Self Expanding Stent The Zilver Vascular Stent is intended recall — FDA Z-2309-2012

Cook Ireland received a product complaint from Japan on December 22nd, 2011. The complainant reported that a 5 French Zilver Vascular Stent (part number…

Class IIclassification
Z-2309-2012official record ID
Sep 12, 2012report date

Recall record at a glance

Official recordZ-2309-2012
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCook Ireland Ltd.
Report dateSep 12, 2012
Recall initiation dateJan 5, 2012
Quantity15 units of ZIV5-18-125-8-80 LOT# CF694429
LocationLimerick, Peroc, N/A, Ireland

Official product description

Zilver 518 Vascular Self Expanding Stent The Zilver Vascular Stent is intended for use as an adjunct to percutaneous transluminal angioplasty (PTA) in the treatment of symptomatic disease of iliac arteries up to 100 mm in length with a reference vessel diameter of 5 to 9 mm. Patients should be suitable candidates for PTA and / or stent treatment.

Why the product was recalled

Cook Ireland received a product complaint from Japan on December 22nd, 2011. The complainant reported that a 5 French Zilver Vascular Stent (part number ZIV5-18-125-8-80 from Lot# CF694429) was inserted into a sheath over a wire guide where it became lodged in the sheath hub. After further examination, the user determined that the product was a 6 French Zilver Vascular Stent (part number ZIV6-35

Distribution information

Worldwide distribution: USA (nationwide) and country of: Japan.

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Other.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.