Recall record at a glance
| Official record | Z-0080-2017 |
|---|---|
| Category | Medical Device Recalls |
| Classification | Class II |
| FDA record status | Terminated |
| Recalling firm | Contour Fabricators Inc. |
| Report date | Oct 19, 2016 |
| Recall initiation date | Sep 9, 2016 |
| Quantity | 447 |
| Location | Fenton, MI, United States |
Official product description
Zero-Gravity Radiation Protection System Radiology Personnel protective shield.
Why the product was recalled
Two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition may cause the cable to break suddenly and the body shield to fall. While no injuries or deaths occurred in either failure, the Zero-GravityTM body shield weighs 55 pounds, and in the event of a cable failure, the body shield will fall immediately. A fall of this nature places the user, associated staff and the patient at risk for a potential injury.
Distribution information
Nationwide. Canada. AUSTRALIA AUSTRIA BELGIUM BRUNEI CHINA FRANCE GERMANY INDA INDONESIA ISRAEL ITALY KSA MALAYSIA NETHERLANDS POLAND QATAR QUEBEC SAUDI ARABIA SINGAPORE SOUTH AFRICA TAIWAN (R.O.C.)
How to read the classification
FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.
Additional record details
Recall type: Voluntary: Firm initiated. Initial notification: Letter.
What to do next
If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.
About this record
Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.
FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.