FDA recall record

Cook Zenith Alpha Abdominal Endovascular Graft - Distributed OUS only recall — FDA Z-2964-2020

May contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or…

Class IIclassification
Z-2964-2020official record ID
Sep 23, 2020report date

Recall record at a glance

Official recordZ-2964-2020
CategoryMedical Device Recalls
ClassificationClass II
FDA record statusTerminated
Recalling firmCook Inc.
Report dateSep 23, 2020
Recall initiation dateAug 11, 2020
Quantity684 units
LocationBloomington, IN, United States

Official product description

Zenith Alpha Abdominal Endovascular Graft - Distributed OUS only. Indicated for the endovascular treatment of patients with abdominal aortic or aorto-iliac aneurysms Reference Part Number (RPN): ZIMB-22-108 ZIMB-22-118 ZIMB-22-128 ZIMB-22-70 ZIMB-22-84 ZIMB-22-98 ZIMB-24-108 ZIMB-24-118 ZIMB-24-128 ZIMB-24-70 ZIMB-24-84 ZIMB-24-98 ZIMB-26-108 ZIMB-26-118 ZIMB-26-128 ZIMB-26-70 ZIMB-26-84 ZIMB-26-98 ZIMB-28-108 ZIMB-28-118 ZIMB-28-128 ZIMB-28-70 ZIMB-28-84 ZIMB-28-98 ZIMB-30-108 ZIMB-30-118 ZIMB-30-128 ZIMB-30-70 ZIMB-30-84 ZIMB-30-98 ZIMB-32-108 ZIMB-32-108 ZIMB-32-118 ZIMB-32-128 ZIMB-32-70 ZIMB-32-84 ZIMB-32-98 ZIMB-36-108 ZIMB-36-118 ZIMB-36-128 ZIMB-36-84 ZIMB-36-98

Why the product was recalled

May contain excess glue within the handle of the delivery system, resulting in an inability to fully deploy the graft via the standard method or troubleshooting deployment method provided in the IFU

Distribution information

Foreign Only: Canada SWEDEN AUSTRIA Belgium CZECH REPUBLIC DENMARK ESPA¿A FINLAND France GERMANY GREECE HUNGARY IRELAND ISRAEL Italia Nederland NORWAY POLAND PORTUGAL REUNION SERBIA SLOVENIA SOUTH AFRICA SWEDEN UNITED ARAB EMIRATES UNITED KINGDOM AUSTRALIA Hong Kong JAPAN NEW CALEDONIA New Zealand Thailand

How to read the classification

FDA Class II indicates a situation in which use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or where the probability of serious consequences is remote. The classification shown here comes from the FDA record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: E-Mail.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.