FDA recall record

Cook Medical Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part recall — FDA Z-0043-2026

Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization.

Class Iclassification
Z-0043-2026official record ID
Oct 22, 2025report date

Recall record at a glance

Official recordZ-0043-2026
CategoryMedical Device Recalls
ClassificationClass I
FDA record statusOngoing
Recalling firmCook Medical Inc.
Report dateOct 22, 2025
Recall initiation dateSep 18, 2025
Quantity695
LocationBloomington, IN, United States

Official product description

Zenith Alpha 2 Thoracic Endovascular Graft, Catalog Prefix ZTA2- Reference Part Numbers ZTA2-P-40-117-W ZTA2-P-40-167 ZTA2-P-40-167-W ZTA2-P-40-217-W ZTA2-P-42-121 ZTA2-P-42-121-W ZTA2-P-42-173 ZTA2-P-42-173-W ZTA2-P-42-225 ZTA2-P-42-225-W ZTA2-P-44-125 ZTA2-P-44-125-W ZTA2-P-44-179 ZTA2-P-44-179-W ZTA2-P-44-233 ZTA2-P-44-233-W ZTA2-P-46-125 ZTA2-P-46-125-W ZTA2-P-46-179 ZTA2-P-46-179-W ZTA2-P-46-233-W ZTA2-PT-40-36-167 ZTA2-PT-40-36-167-W ZTA2-PT-40-36-217 ZTA2-PT-40-36-217-W ZTA2-PT-42-38-173 ZTA2-PT-42-38-173-W ZTA2-PT-42-38-225 ZTA2-PT-42-38-225-W ZTA2-PT-44-40-179 ZTA2-PT-44-40-179-W ZTA2-PT-44-40-233-W ZTA2-PT-46-42-179 ZTA2-PT-46-42-179-W ZTA2-PT-46-42-233 ZTA2-PT-46-42-233-W ZTA2-P-46-233 ZTA2-PT-44-40-233 ZTA2-P-40-217 ZTA2-P-40-117 ZTA2-P-40-167-DEMO ZTA2-P-44-152-W ZTA2-P-40-142 ZTA2-P-44-152 ZTA2-P-40-142-W

Why the product was recalled

Affected devices may contain PTFE coating scrapings. Scrapings could be released during device deployment, potentially causing intravascular embolization.

Distribution information

Domestic distribution to AL AR AZ CA CT DC DE FL GA IL IN LA MA MD MI MN MO NC NJ NY OH OR PA SC TN TX UT VA WA WI WV International distribution to Switzerland, Germany, Italy, Poland, and Sweden

How to read the classification

FDA Class I indicates a situation in which there is a reasonable probability that use of or exposure to the product will cause serious adverse health consequences or death. The classification shown here comes directly from the FDA enforcement record.

Additional record details

Recall type: Voluntary: Firm initiated. Initial notification: Letter.

What to do next

If you believe you have the recalled product, use the recall number and firm information above to confirm the official notice and follow the instructions issued by the FDA, recalling firm, retailer or product manufacturer. For drugs and medical devices, decisions about treatment or medical care should be discussed with a qualified health professional.

About this record

Recall Lookup USA preserves the most complete source information successfully retrieved. We do not create missing facts or infer affected lots or VINs. The official record may still change, so use the identifier above when verifying current instructions.

FDA source limitation: openFDA enforcement reports are a reference dataset, not a live public-alert or recall-lifecycle service. FDA notes that enforcement-report status may not be updated after classification. Confirm current actions with the FDA, recalling firm or appropriate healthcare professional when relevant.

Related recall records

Other records from the same recalling company are shown first when available.